Vortex Measurement Principle

Vortex FX300

SITRANS F X vortex flowmeters measure flow rate by detecting the frequency at which alternating vortices are shed from a bluff body inserted into the flow stream. This principle of measurement is known as Von Karman’s vortex street  rinciple: alternating vortices form behind an object in a stream. The frequency of the alternating vortices is proportional to the flow rate. The passage of a vortex causes a slight stress on a pick-up placed downstream of the bluff body. The stress is picked up and counted as pressure surges by a dual Piezo crystal placed inside the wing.
SITRANS F X vortex flowmeters provide accurate volumetric and mass flow measurement of steam, gases and liquids as an all-in-one solution with integrated temperature and pressure compensation.
The SITRANS FX300 is a flowmeter in a single or dual transmitter version, suitable for measuring industrial steam, gases, as well as conductive and non-conductive liquids, e.g. steam (saturated steam, superheated steam), industrial  ases (compressed air, nitrogen, liquefied gases, flue gases), and conductive and non-conductive liquids (demineralized water, boiler feed water, solvents, heat transfer oil).
The unit is available in a compact or a remote version with up to 30 meter distance from flowmeter to transmitter. When ordering a remote version the transducer cables are pre-mounted and ready for installation.

Applications:
  • industrial steam
  • gases
  • conductive and non-conductive fluids
Download:
Vortex FX300
Inquiry Sheet
  • All devices have 2-wire technology and HART communication
  • Temperature compensation for saturated steam as standard feature
  • Integrated temperature and pressure measurement enabling direct compensation of density
  • Pressure, temperature and flow can be read at a single point. No additional installation of pressure and temperature sensors
  • Direct measurement of energy or energy consumption
  • Optimum process reliability thanks to Intelligent Signal Processing (ISP) - stable readings, free of external perturbations
  • Fully welded stainless steel construction with high corrosion, pressure and temperature resistance
  • Maintenance-free sensor design
  • Ready to use due to plug & play feature. No additional cabling work
  • Minimal pressure drop
  • Compact or remote design
  • Pressure and temperature compensation for fluctuating volume flows
  • Measurement of consumption in compressed air systems
  • No risk of deposits or damage (sensor in the turbulent area)
  • All units parameterized prior to delivery